INTUITIVE SURGICAL, INC NONE; ENDOSCOPE PLUS Back to Search Results
Model Number 470057-05
Device Problem Image Orientation Incorrect (1305)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/30/2023
Event Type  malfunction  
Manufacturer Narrative
Based on the claim against the product by the customer noting an image mirroring issue, an investigation was completed to determine the cause of this reported event.Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The endoscope was analyzed and found to have endoscope adapter (aea) binding or friction issue.The profile tube was damaged, and contamination/corrosion was observed.The shaft was found to have mechanical damage.There was discoloration of housing and transmittance test failure.The complaint regarding image orientation issue was confirmed by failure analysis, which indicates that the device did contribute to the customer reported issue.This complaint is considered a reportable event due to the following conclusion: it was alleged that the endoscope image was mirrored.Poor camera control could result in unintuitive motion and subsequent tissue damage.At this time, the root cause of the failure is unknown.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur.
 
Event Description
It was reported that during a da vinci-assisted radical cystectomy with neobladder surgical procedure, the image from the 30-degree endoscope was mirrored.The procedure was completing as planned with no reported injury.Intuitive surgical inc.(isi) followed up with the initial reporter and obtained the following additional information: during the operation the instruments suddenly could no longer be controlled, and the camera image appeared double, and mirror inverted.After contacting the technician, the optics were replaced, after which all faults were rectified, and the operation was completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NONE
Type of Device
ENDOSCOPE PLUS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key16781276
MDR Text Key313692330
Report Number2955842-2023-12093
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier00886874116555
UDI-Public(01)00886874116555
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K191736
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial
Report Date 03/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number470057-05
Device Catalogue Number470057
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/30/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES.
-
-