• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. CEMENTED CR FEMORAL SZ 3; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. CEMENTED CR FEMORAL SZ 3; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED Back to Search Results
Model Number 200-01-03
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Synovitis (2094)
Date of Event 07/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending investigation.D10: 1109088, 200-23-11 - cr tibial insert sz3, 11mm; 2731737, 200-02-38 - three peg patella 38mm; 2875528, 204-04-33 - trapezoid tibial tray sz 3f/3t; 2942120, 204-32-01 - fluted stem extension 11l x 12 mm; 3581467, 204-70-00 - tibial stem ext.Screw; aa6861, 1400-ag - cemex gento low viscosity 40g kit.H7: z-0019-2022.
 
Event or Problem Description
As reported by legal notification, the patient had an initial left tka on (b)(6) 2014.The patient was revised on (b)(6) 2022.Joint fluid was obtained and sent for culture and sensitivity x3.Extensive synovitis from the polyethylene was noted.The femoral component was grossly loose and was removed with osteotomes.The tibial component and patella were also grossly loose and were removed.The patient tolerated the procedure well.No other patient information/medical history reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, femoral loosening, and tibial loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CEMENTED CR FEMORAL SZ 3
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key16782198
Report Number1038671-2023-00782
Device Sequence Number11576308
Product Code JWH
UDI-Device Identifier10885862039538
UDI-Public10885862039538
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K932690
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 01/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/26/2019
Device Model Number200-01-03
Device Catalogue Number200-01-03
Was Device Available for Evaluation? No
Distributor Facility Aware Date03/31/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2023
Supplement Date Received by Manufacturer01/08/2025
Initial Report FDA Received Date04/20/2023
Supplement Report FDA Received Date01/09/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/28/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexUnknown
-
-