| Model Number |
24674 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in a shunt in a severely tortuous and severely calcified vein.A 7.0 x 40, 40cm mustang balloon catheter was advanced for dilation.However, during the third inflation at 10-20 atmospheres for 30 seconds, the balloon ruptured.The device was removed, and the procedure was completed with another of same device.No patient complications were reported, and the patient was in good condition post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Iinitial reporter facility name: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
(e1) initial reporter facility name: (b)(6) clinic.Device evaluated by mfr.: mustang 7.0 x 40, 40cm was received for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 15mm distal of the proximal markerband.A visual examination observed no issues or damage to the tip of the device.A visual and tactile examination found no kinks or damage the shaft of the device.A visual examination observed no issues with the markerbands of the device.Both markerbands were undamaged and present on the shaft of the device.This concludes the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in a shunt in a severely tortuous and severely calcified vein.A 7.0 x 40, 40cm mustang balloon catheter was advanced for dilation.However, during the third inflation at 10-20 atmospheres for 30 seconds, the balloon ruptured.The device was removed, and the procedure was completed with another of same device.No patient complications were reported, and the patient was in good condition post procedure.
|
| |
|