ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, FOOTSWITCH; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065751762
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 03/23/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A physician reported that during cataract while phacoemulsification of the third segment the foot pedal lost connection with the ophthalmic console the physician tried to reconnect the foot pedal and asked the staff for main cord, cored was not located quick enough to complete the procedure which resulted choroidal bleeding.The procedure was not completed.Eye was sutured, diuretic and topical medications were used to stabilize the eye.Patient to return on another day to have procedure completed.
 
Additional Manufacturer Narrative
Specific product identifiers (serial number) were not provided and could not be determined at this time.However, all device history records are reviewed prior to product release to ensure the product was manufactured in compliance with the device master record and meets release criteria.A review for complaints reported against this serial number cannot be performed as the lot/batch/serial number is unknown.The serial is unknown.Therefore, a service history review cannot be performed.The product under investigation is not a serviceable device.Therefore, a service record review was not performed.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate the manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, FOOTSWITCH
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16784877
Report Number2028159-2023-00582
Device Sequence Number5691503
Product Code HQC
UDI-Device Identifier00380657517626
UDI-Public00380657517626
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K121555
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number8065751762
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/27/2023
Supplement Date Received by Manufacturer06/23/2023
Initial Report FDA Received Date04/21/2023
Supplement Report FDA Received Date07/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CENTURION VISION SYSTEM
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age86 YR
Patient SexMale
-
-