| Model Number |
N/A |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/28/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: canada.H3: customer has indicated that the product will not be returned because requested but not returned by hospital.Once the investigation has been completed, a follow-up mdr will be submitted.H3 other text : not returned by hospital.
|
| |
|
Event or Problem Description
|
|
It was reported that the device was discovered to have missing components.No more information is available.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event was confirmed by review of photo provided.Visual evaluation of the provided device photographs shows signs of repeated use and have one or both of components disassembled / missing the components.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.H6: type of investigation - device not returned and communication/interviews is n/a which was reported on initial report.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned device shows sign of use with some wear and both instrument found to be missing one of two ball bearings.Additional information does not affect the root cause.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|