| Model Number |
M00545180 |
| Medical Device Problem Codes |
Break (1069); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 44 was used in the papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2023.During the procedure, the cutting wire of the autotome rx 44 broke when they started the papillotomy.It was reported that no part of the device was detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event and the patient's condition at the conclusion of the procedure was reported to be stable.Note: a photo of the complaint device outside the patient was provided by the customer and showed the cutting wire was broken and kinked.
|
| |
|
Additional Manufacturer Narrative
|
|
Imdrf device code a0401 captures the reportable event of cutting wire broken.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h10: the returned autotome rx 44 was analyzed, and a visual evaluation noted that the cutting wire was broken from the proximal pierced hole, and also it was kinked, which are consistent with the findings when the device was observed under magnification and per media inspection in the provided photo.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was broken, and kinked from the proximal pierce hole.Based on the condition of the device, the problem found could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire's integrity and cause a premature cutting wire fatigue.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can kink the cutting wire.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.Block h11 (correction): block h6 (device codes) has been corrected.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that an autotome rx 44 was used in the papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6), 2023.During the procedure, the cutting wire of the autotome rx 44 broke when they started the papillotomy.It was reported that no part of the device was detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event and the patient's condition at the conclusion of the procedure was reported to be stable.Note: a photo of the complaint device outside the patient was provided by the customer and showed the cutting wire was broken and kinked.
|
| |
|