| Model Number |
21-2120-0105-231 |
| Medical Device Problem Code |
Product Quality Problem (1506)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the cadd pump air in line detector was bad and needed to be replaced.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: additional information is provided for h.2, h.3, and h.6.One device was received for evaluation.Visual inspection revealed the device had no physical damage.Functional testing was performed, and the reported issue was able to be replicated and confirmed.The root cause was determined to be air in the line.Service history review identified the complaint was not related to a previous service of the device within the review period.For all enquiries or follow-up questions related to the record, do not use regulatory.Responses@smiths-medical.Com located in sections g.1., please direct those to the following: regulatory.Responses@icumed.Com.
|
| |
|