|
An olympus representative reported to olympus on behalf of the customer that during open oesophagectomy using this ultrasonic surgical device, a circuit error was observed however, the device was still working fine.When the user was grasping a tissue, one of the probe tips fell to the abdominal cavity.The probe was retrieved with non-toothed forceps.The retrieved probe was checked and it was all intact and completely removed from patient's cavity.There were no reports of further patient or user harm associated with this event.
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was evaluated by olympus, and the customer¿s reported complaint (broken probe) was confirmed.In addition, the following non-reportable findings were observed: scratches were observed around the broken area.The coating of the grasping section was partially peeled off.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been more than 1 year since the subject device was manufactured.Based on the results of the investigation, the root cause could not be determined.However, the broken probe was likely caused by one of the following mechanisms: 1) during the output activation in seal & cut mode, the probe was in contact with hard tissue, metal objects or surgical instruments.2) scratches were generated on the probe.3) a force to activate the output in seal & cut mode, or a force to grasp the tissue was applied to the probe.Therefore, cracks were generated at the scratched area and the error occurred.4) a force was applied to the probe causing it to break.The event can be detected/prevented by following the instructions for use (ifu) which state: ¿the thunderbeat instrument should be used for soft tissue.Do not activate output while grasping hard tissue such as bone or highly calcified tissue, or hard objects such as metal clips, stapler, other instruments, or forceps, and others.Otherwise, it may cause the probe tip to be scratched or come into direct contact with the metal area of the grasping section as the heat generated by the friction between the hard object and the probe tip could cause wear/deforming/split/protruding/partial separating of the tissue pad.In turn, the probe tip may break before displaying an error window or generating an alarm tone.¿ ¿during output, do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments.Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities, which may lead to burns of the tissue by a high-frequency leak current output due to destruction of the insulation structure.¿ ¿if the grasping section, metal-exposed area around it or the probe tip gets stuck to tissue during treatment, wipe it with a soft object such as a piece of gauze or a brush.Do not attempt to scrape it with a sharp object such as a scalpel or the tip of tweezers.Otherwise, the grasping section, metal-exposed area around it, the fluorine resin part, a coated surface or the probe tip may be scratched and damaged, which may lead to fall-off of the damaged part into the body cavity or burns of the tissue by a high-frequency leak current output due to destruction of the insulation structure.¿ this supplemental report includes a correction to the following fields: d4, d9, g2, and h3.Also, new information has been added to h4.Olympus will continue to monitor field performance for this device.
|