| Model Number |
0620040600 |
| Medical Device Problem Codes |
No Flow (2991); Pressure Problem (3012)
|
| Health Effect - Clinical Codes |
Peritoneal Laceration(s) (2003); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the case was delayed a few days due to loss of insufflation during procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that the case was delayed a few days due to loss of insufflation during procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: cib, levian, rep, calibration or something wrong with the flow, wcb.*update: loss of insufflation for a little during a case until worked for a while and then stopped working properly after case was over.The failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root cause could be third party repair.The unit is also past due for calibration.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|