| Model Number |
0502503030 |
| Medical Device Problem Code |
Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was blurry image during procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Additional Manufacturer Narrative
|
|
This scope was not received for evaluation at stryker endoscopy.This scope was received at henke for evaluation.Based on the henke , the reported failure "broken" was confirmed.According to henke: scope evaluated as a level 5: distal tip and fiber damage, particulate under distal negative, scratches on distal tip negative, loose prism and broken lenses in system.Confirmed third party repair: due to non-oem unknown needle weld, unknown tip work and sandblasted finish.Probable root cause for this failure is: the complaint for ¿broken¿ has been confirmed and is due to customer used and handling.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was blurry image during procedure.
|
| |
|
Search Alerts/Recalls
|