Model Number 135-5114
Medical Device Problem Codes Off-Label Use (1494); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 03/24/2023
Type of Reportable Event Malfunction
Event or Problem Description
Biopoly learned from physician about osteolysis/bone resorption in one of his patients, and that he removed the biopoly implant.
 
Additional Manufacturer Narrative
Biopoly received the explant for inspection.The explant was examined under magnification and appeared to show signs of abrasive wear.The remaining biopoly material was securely attached to the titanium stem and the osteosync material was also securely attached.The implant had been secured in the bone via press fit; however, there was no evidence that bone cement had been used.The implant manufacturing lot met all specifications.Although it is not confirmable, the cause appears to have been overstuffing of the joint along with the phalanx being devoid of cartilage, a contraindication, leading to excessive and abrasive biopoly wear and leading to osteolysis.Wear and damage of the implant are possible adverse effects listed in the instructions for use (ifu) document and considered risks in the risk management system for this implant.Many possible factors could lead to damage to the implant during implantation, including failure to follow instructions for use, use of implant against non-cartilage surface, or improper patient selection (age, bone quality, cartilage health).If additional information is received, information will be reviewed for reportability and submitted for follow-up as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOPOLY GREAT TOE HEMIARTHROPLASTY IMPLANT
Common Device Name
TOE PROSTHESIS
Manufacturer (Section D)
BIOPOLY LLC
7136 gettysburg pike
fort wayne IN 46804
Manufacturer (Section G)
BIOPOLY LLC
7136 gettysburg pike
fort wayne IN 46804
Manufacturer Contact
stone miguel
7136 gettysburg pike
fort wayne, IN 46804
2609996135
MDR Report Key16794259
Report Number3008108789-2023-00002
Device Sequence Number20555117
Product Code KWD
UDI-Device Identifier00850012119388
UDI-Public(01)00850012119388(17)260520(10)2133827
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K203634
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/20/2026
Device Model Number135-5114
Device Catalogue Number135-5114
Device Lot Number2133827
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2023
Initial Report FDA Received Date04/24/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
Patient SexFemale
-
-