| Model Number |
900-0632-00 |
| Medical Device Problem Codes |
Patient-Device Incompatibility (2682); Biocompatibility (2886)
|
| Health Effect - Clinical Codes |
Blister (4537); Skin Inflammation/ Irritation (4545)
|
| Date of Event |
03/09/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient called stating patch has caused reaction - sore painful peeling skin.Noticed since day 1.The patient's skin was prepped with alcohol and patch replaced daily.Neosporin cream was used on skin.The patient is still currently wearing device.Skin sensitivity known prior to receiving device.Patient services called the patient 3 times obtain additional information and no answer.Additional information was provided on 12 april 2023 from patient services from the patient: the patient reported that they were prescribed hydrocortisone cream by their doctor.The patient could not recall the name of brand that was prescribed.
|
| |
|
Additional Manufacturer Narrative
|
|
Engineering evaluation was unable to be performed as the electrode/device was not returned.Allegation is confirmed through images of patient skin irritation and is most probable to be a bio-incompatibility issue with the electrode adhesive.Marsi, skin burn, and associated symptoms may inherently occur under the course of ecg monitoring.No single factor or combination of factors can be attributable to electrode skin irritation and associated symptoms.The product labeling advises patients of alternate options and other steps to take if skin irritation develops, including healthcare professional contact as needed.
|
| |
|