ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number TABLETOP
Device Problem Decrease in Pressure (1490)
Patient Problems Hemorrhage/Bleeding (1888); Intraocular Pressure Decreased (4468)
Event Date 02/17/2023
Event Type  Injury  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that an ophthalmic console stopped maintaining appropriate eye pressure in the left eye causing a loss of pressure and loss of chamber inflation and caused a small subsequent hemorrhage during surgery and the patient's symptoms were resolved.
 
Manufacturer Narrative
The company representative did not confirm nor replicate the reported event.The system was tested and found to meet product specifications.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A non-conformance based review of the serial number was performed and a potential contributing factor to the reported complaint was identified.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.The system was found to have no problem.Therefore, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16795003
MDR Text Key313817018
Report Number2028159-2023-00587
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657511501
UDI-Public00380657511501
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 05/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/17/2012
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age70 YR
Patient SexFemale
-
-