• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORPORATION TURBOVAC 90 XL ICW; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORPORATION TURBOVAC 90 XL ICW; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number ASC1336-01
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Loss of Range of Motion (2032); Unspecified Tissue Injury (4559)
Date of Event 03/24/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that on the pmcf "pmcf for turbovac 90 xl icw, multivac 50 xl icw & topaz ez ifs", was found that during a procedure using a turbovac 90 xl icw wand there was an unwanted tissue damage and bleeding; which led to a post operative bleeding and stiffnes in the joint.It is unknown how the events were treated.Patients¿ outcome is unknown.No further information is available.
 
Additional Manufacturer Narrative
H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURBOVAC 90 XL ICW
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORPORATION
7000 west william cannon drive
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16796949
Report Number3006524618-2023-00161
Device Sequence Number3251543
Product Code GEI
UDI-Device Identifier00817470006520
UDI-Public00817470006520
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K033584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberASC1336-01
Device Catalogue NumberASC1336-01
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2023
Supplement Date Received by Manufacturer07/07/2023
Initial Report FDA Received Date04/24/2023
Supplement Report FDA Received Date07/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age55 YR
Patient SexMale
-
-