| Model Number |
72201775 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a procedure the biosure ha broke while attempting to fix the tendon to the bone.The procedure was completed without surgical delay using a smith and nephew back up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: case (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5 and h6 (health effect - impact code) updated.
|
| |
|
Event or Problem Description
|
|
It was reported that during an acl procedure, the biosure ha broke while attempting to fix the tendon to the bone.All the broken pieces were removed from the patient using tweezers.The procedure was completed without surgical delay using a smith and nephew back up device.The back-up device was used in the originally drilled bone hole.No further complications were reported.Patient current status is good.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference (b)(4).H6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Search Alerts/Recalls
|