MAUDE Adverse Event Report: BD MEDICAL - DIABETES CARE BD ULTRA-FINE¿ II SHORT NEEDLE INSULIN SYRINGE; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL - DIABETES CARE BD ULTRA-FINE¿ II SHORT NEEDLE INSULIN SYRINGE; PISTON SYRINGE Back to Search Results
Catalog Number 326725
Device Problem Missing Information (4053)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2023
Event Type  malfunction  
Event Description
It was reported that 1700 bd ultra-fine¿ ii short needle insulin syringe was missing label information.The following information was provided by the initial reporter: lack information, information not clear, deformed & torn (b)(6) 2023: rcc reviewed the supporting evidence (photo) and the box packages were deformed and torn.
 
Manufacturer Narrative
H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Manufacturer Narrative
H.6.Investigation summary: no physical samples were received; the investigation was performed based on the photo provided.This is the 2nd complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Embecta was able to confirm the customer indicated failure.The reported issue was observed and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
 
Event Description
It was reported that 1700 bd ultra-fine¿ ii short needle insulin syringe was missing label information.The following information was provided by the initial reporter: lack information, information not clear, deformed & torn.On 06-apr-2023: rcc reviewed the supporting evidence (photo) and the box packages were deformed and torn.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ULTRA-FINE¿ II SHORT NEEDLE INSULIN SYRINGE
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
phillip emmert
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key16802044
MDR Text Key313917324
Report Number1920898-2023-00252
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeVM
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number326725
Device Lot Number2137915
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-