Model Number 1650
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/16/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that four (4) of the patient batteries were not supporting >3-4 hours of charge.The batteries were removed from ser vice.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Investigation of this event is pending and a supplemental report will be sent upon its completion.Additional products: d1: heartware ventricular assist system ¿battery model#: 1650 / catalog#: 1650/ expiration date: 31-aug-2019 / serial#:(b)(6) udi#: (b)(4) mfg date: 05-aug-2018 patient ime code(s): e2403 h6: imf code(s): f26: img code(s): g02002 fda device code(s): a0705 fda method code(s): b21 fda results code(s): c21 fda conclusion code(s): d16 heartware ventricular assist system ¿battery model#: 1650 / catalog#: 1650/ expiration date: 31-aug-2019 / serial#: (b)(6).Udi#: (b)(4).Mfg date: 05-aug-2018 patient ime code(s): e2403 imf code(s): f26 img code(s): g02002 fda device code(s): a0705 fda method code(s): b21 fda results code(s): c21 fda conclusion code(s): d16 heartware ventricular assist system ¿battery model#: 1650 / catalog#: 1650/ expiration date: 31-aug-2019 / serial#: (b)(6).Udi#: (b)(4).Mfg date:05-aug-2018 patient ime code(s): e2403 imf code(s): f26 img code(s): g02002 fda device code(s): a0705 fda method code(s): b21 fda results code(s): c21 fda conclusion code(s): d16### medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
A supplemental report is being submitted for investigation completion.Product event summary: four (4) batteries (b)(6) were not returned for evaluation.Review of the manufacturing documentation confirmed that the associated devices met all requirements for release.Log file analysis was not performed since log files were not available for analysis.As a result, the reported event could not be confirmed due to insufficient evidence.Applicable risk documentation and experience with events of similar circumstances were considered; events involving batteries that rapidly discharge can be attributed but not limited to soldering or welding defects.Additional products: d4: serial or lot#: (b)(6) h6: fda method code(s): b14, b17 h6: fda results code(s): c19 h6: fda conclusion code(s): d14 d4: serial or lot#: (b)(6) h6: fda method code(s): b14, b17 h6: fda results code(s): c19 h6: fda conclusion code(s): d14 d4: serial or lot#: (b)(6) h6: fda method code(s): b14, b17 h6: fda results code(s): c19 h6: fda conclusion code(s): d14 investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer (Section G)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key16807302
Report Number3007042319-2023-02044
Device Sequence Number12288513
Product Code DSQ
UDI-Device Identifier00888707000369
UDI-Public00888707000369
Combination Product (Y/N)N
Initial Reporter CountryVM
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/31/2019
Device Model Number1650
Device Catalogue Number1650
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2023
Supplement Date Received by Manufacturer06/12/2023
Initial Report FDA Received Date04/25/2023
Supplement Report FDA Received Date06/14/2023
Date Device Manufactured08/05/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MCS UNKNOWN VAD
Patient SexMale
-
-