• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE Back to Search Results
Model Number LTF-S190-10
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer returned the endoeye flex deflectable videoscope because it displayed an error code e216 (scope communication error) when the up/down angulation control was used.It is unknown when the event occurred.No patient injury reported.
 
Manufacturer Narrative
The device was returned to olympus medical for evaluation.During testing, the reported error e216 was confirmed.In addition, there was no image on the monitor when scope was positioned at certain angles.At this time, the investigation has determined the issues may have occurred due to damage to the image sensor unit or mounting components on the electrical board.In addition, the following non-reportable malfunctions were found during the device evaluation: insufficient angle, bending section scratches, bending section adhesion chip, image guide snake tube backlash, video connector deformation, video connector damaged.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the suggested events could not be concluded, although it can be presumed that the events occurred due to the damage of image sensor unit (disconnection, etc.) or breakage of the mounting components (ic chip, capacitor, etc.) of the electric board due to use stress, external factors, or handling.The events can be detected/prevented by following the inspection method for the event is described as follows in ¿chapter 3 preparation and inspection, section 3.8 inspection of the endoscopic system¿ as below."[inspection of the endoscopic image].Confirm that the wli and nbi endoscopic images are normal.1 before inspection, wipe the objective lens using clean lint-free cloths moistened with saline solution or sterilized water.2 observe the palm of your hand in the wli and nbi endoscopic images.3 confirm that light is output from the endoscope¿s distal end.(see figure 3.23).4 adjust the brightness level as appropriate.5 confirm that the wli and nbi endoscopic images are free from noise, blur, fog, or other irregularities.6 turn the angulation control levers slowly in each direction until it stops 7 confirm that the wli and nbi endoscopic images do not momentarily disappear or display any other irregularities." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE
Type of Device
FLEX DEFLECTABLE VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16808748
MDR Text Key314113609
Report Number9610595-2023-06660
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier04953170310386
UDI-Public04953170310386
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K111425
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLTF-S190-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/26/2023
Initial Date FDA Received04/25/2023
Supplement Dates Manufacturer Received04/24/2023
Supplement Dates FDA Received05/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/11/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-