• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941832400
Medical Device Problem Codes Entrapment of Device (1212); Difficult to Remove (1528); Failure to Deflate (4060)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/06/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that deflation issues occurred.A 12mm x 2.50mm nc quantum apex and a 4.00 x 32mm synergy xd were selected for use.During the procedure, after the stent was successfully delivered and expanded, it was noted that the balloon got stuck in the guide.Another inflation device was used in an attempt to fully deflate the balloon, but it remained stuck in the guide, so a wire was used to pop and further deflate the balloon.Next the nc balloon got stuck in the guide after deflation and the nc balloon was also popped to further deflate.The balloon was removed and the procedure was completed successfully.There were no patient complications.
 
Event or Problem Description
It was reported that deflation issues occurred.A 12mm x 2.50mm nc quantum apex and a 4.00 x 32mm synergy xd were selected for use.During the procedure, after the stent was successfully delivered and expanded, it was noted that the balloon got stuck in the guide.Another inflation device was used in an attempt to fully deflate the balloon, but it remained stuck in the guide, so a wire was used to pop and further deflate the balloon.Next the nc balloon got stuck in the guide after deflation and the nc balloon was also popped to further deflate.The balloon was removed and the procedure was completed successfully.There were no patient complications.
 
Additional Manufacturer Narrative
H6 device codes: corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jeff wallner
4100 hamline ave n
arden hills, MN 55112
6515811560
MDR Report Key16809412
Report Number2124215-2023-19072
Device Sequence Number2996991
Product Code NIQ
UDI-Device Identifier08714729981169
UDI-Public08714729981169
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/07/2024
Device Model NumberH7493941832400
Device Lot Number0029995432
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/06/2023
Supplement Date Received by Manufacturer05/05/2023
Initial Report FDA Received Date04/25/2023
Supplement Report FDA Received Date05/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/08/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-