ABBOTT GMBH ALINITY I HIV AG/AB COMBO REAGENT KIT; TEST, HIV DETECTION Back to Search Results
Model Number 08P0721
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event or Problem Description
The customer observed a false nonreactive alinity i hiv ag/ab combo result for one patient.The alinity i hiv ag/ab combo result was nonreactive for sid (b)(6), and a viral load was also ordered.The viral load result was ¿detected, 60000 copies/ml¿.The doctor then asked to retest the hiv sample on the alinity and the retest result was also nonreactive.The patient had another specimen collected 2 days later and resulted 7.55 s/co.No impact to patient management was reported.
 
Additional Manufacturer Narrative
Data and information provided by the customer were reviewed and support the complaint issue without indication for any additional issue.The ticket search by lot indicates that the reagent lot performs as expected for this product.Device history record review was performed on lot 48132be00, which did not show any potential non-conformances, deviations, or non-conformances.Trending review did not identify any trends.In-house testing of a retained reagent kit of the complaint lot was performed, showing that the lot generates the expected results.In addition, the clinical sensitivity of the lot was evaluated by testing two commercially available seroconversion panels (zeptometrix hiv 9013 and hiv 9016).The seroconversion panel results were compared to architect hiv test results provided by zeptometrix.The lot detected the same bleeds as reactive for the seroconversion panels.Based on these data it was shown that the sensitivity performance of the complaint lot is not adversely affected.Note: alinity i hiv ag/ab combo and architect hiv ag/ab combo utilize the same reagents and sample/reagent ratios.The returned specimen sample id (b)(6) was tested with a retained kit of alinity i hiv ag/ab combo (us) reagent lot 48132be00 and a non-reactive result (0.73 s/co) was obtained, confirming the customer¿s observation.The specimens was further tested with western blot mikrogen recomline hiv-1 & hiv-2 igg and showed a negative result; no bands were identified.Innotest hiv antigen mab testing (detects p24 antigen) provided a negative result (od 0.038, cutoff: 0.050).Note: per complaint information the customer obtained a positive viral load testing result with 60,000 copies/ml.No indication of false non-reactive results was identified, as the alinity i hiv ag/ab combo detects hiv p24 antigen and antibodies to hiv-1 and hiv-2, but not hiv rna.Labeling review concludes that the issue is adequately addressed.Note: in the early phase of an hiv infection there is a delay before detectable levels of antigens and antibodies appear.Therefore, the most likely explanation for the non-reactive test results is that the donors were in the diagnostic window below the detection limit.No malfunction was identified, as in-house testing of return sample showed non-reactive hiv antigen (innotest hiv mab) and antibody (hiv-1/hiv-2 immunoblot) results.Based on the information within the complaint record, the device met performance specifications and performed as intended at the customer site.
 
Event or Problem Description
The customer observed a false nonreactive alinity i hiv ag/ab combo result for one patient.The alinity i hiv ag/ab combo result was nonreactive for sid (b)(6), and a viral load was also ordered.The viral load result was ¿detected, 60000 copies/ml¿.The doctor then asked to retest the hiv sample on the alinity and the retest result was also nonreactive.The patient had another specimen collected 2 days later and resulted 7.55 s/co.No impact to patient management was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY I HIV AG/AB COMBO REAGENT KIT
Common Device Name
TEST, HIV DETECTION
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key16809679
Report Number3002809144-2023-00202
Device Sequence Number14202382
Product Code MZF
UDI-Device Identifier00380740121754
UDI-Public00380740121754
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/18/2023
Device Model Number08P0721
Device Catalogue Number08P07-21
Device Lot Number48132BE00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/06/2023
Supplement Date Received by Manufacturer06/20/2023
Initial Report FDA Received Date04/25/2023
Supplement Report FDA Received Date06/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/29/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY I HIV COMBO 200 T, 08P07-21, 48132BE00; ALNTY I HIV COMBO 200 T, 08P07-21, 48132BE00
Patient SexUnknown
-
-