|
It was reported that during the treatment of an acom aneurysm in remodeling technique, during the first inflation, the balloon was placed at the neck of the aneurysm when the physician noticed air bubbles exiting at the distal tip of the balloon.No air embolism was observed.The scepter xc was exchanged, and the examination continued with good results.No injury was caused by this issue and the patient is doing well.
|
|
Items returned for evaluation: 1 scepter xc.Items not returned for evaluation: n/a.Visual inspection found that the catheter section of the device was kinked proximal to the polyimide ring.The balloon was attempted to be inflated; however, due to the hub's inflation port/lumen being occluded with material consistent with dried contrast, the hub was bypassed to test balloon inflation.With the hub bypassed, the balloon was filled with saline mixture; however, during the inflation it was noticed that two pin hole leaks were present at the proximal end of the balloon.Due to the pinholes causing the balloon to progressively deflate on its own, the presence of air bubbles at the distal tip of the balloon at nominal pressure was unable to be investigated.Per ifu, the following steps and precautions should be taken during preparation to prevent air bubbles within the balloon: balloon preparation: if the balloon is initially inflated with air then maintain constant syringe pressure.Warning: do not fully inflate balloon with air.Maintain balloon inflation below distal balloon marker band.Maintain pressure and do not tilt the balloon until the contrast reaches the distal purge hole and the contrast has completely filled the balloon.Once the balloon has been fully inflated with contrast, inspect balloon for any damages and bubbles.If air bubbles persist, use the aspiration technique outlined in step a otherwise place tip in saline bowl, deflate balloon.Alternate balloon preparation: place balloon catheter into a large bowl of saline for continued hydration.Note: keep the catheter in a horizontal position while submerged in the saline solution.With the distal tip of the balloon catheter submerged in saline, slowly inject contrast/saline solution through the balloon inflation port and allow all the air to escape from the distal air-purge hole.During this air-purging process, caution should be taken to inject slowly to prevent the balloon from inflating with the air.If the balloon is inflated before the contrast/saline solution reaches the balloon, stop air-purging the balloon to allow the balloon to deflate.Slowly inject contrast/saline solution again allowing remaining air to escape through the distal air-purge hole.Reference table 2 for the total prime volume of the inflation hub and the inflation lumen.Additional solution will be required to inflate the balloon.Warning: during air-purging of balloon catheter, inject fluid slowly otherwise balloon rupture may occur.After air-purging the balloon catheter inflation lumen, continue to slowly inflate the balloon with the contrast/saline solution.The balloon will inflate from proximal to distal during the initial inflation.When the balloon catheter is completely inflated, check for the presence of additional air bubbles.If any air bubbles are present, submerge the tip of the balloon catheter in saline and deflate completely, then slowly inflate with the balloon held upright to allow any remaining air to escape from the distal air-purge hole.If bubbles persist, use the aspiration technique outlined in step a, otherwise place tip in saline bowl, deflate balloon and proceed to step b, c or d.Warning: do not deflate the balloon unless the distal tip is submerged in saline or contrast to prevent air from getting back into balloon.Aspiration technique: remove the balloon from the saline bowl and place the distal tip of the balloon in air.Hold vacuum until the inflation port is clear of contrast/saline mix.Close stopcock to inflation lumen, point syringe up and purge air.If no air bubbles are visible inside balloon, place tip in saline bowl, deflate balloon and then proceed to step b, c or d.If air bubbles persist, repeat step a or discard balloon catheter.The investigation of the returned balloon catheter found the catheter kinked proximal to the polyimide ring and the hub occluded with material consistent with dried contrast from the device preparation.After bypassing the hub, the balloon was inflated, and the investigation found two pinhole leaks causing the device to progressively deflate on its own.Due to the pinholes causing the balloon to progressively deflate on its own, the presence of air bubbles at the distal tip of the balloon when at nominal pressure was unable to be investigated.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.
|