| Catalog Number |
56538CB2/B |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/18/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow-up report will be submitted once the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a percutaneous transluminal angioplasty [pta] procedure within the patient's periphery, a catheter tip detached.The physician had acquired right retrograde popliteal artery access and had negotiated the patient's contralateral left ilio-femoral conduit with an interventional guidewire and a 5f catheter.During catheter manipulations under fluoroscopy the catheter tip detached but remained on the guidewire.The detached catheter tip was removed along with the guidewire.No additional medical intervention was deemed necessary.A new device was used to complete this procedure.No additional patient consequences to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect medical device has been returned for evaluation.The device was visually and microscopically evaluated.The complaint is confirmed.The root cause can be contributed to the manufacturing process.The device history record was reviewed, and no exception documents were found.A search of the complaint database was performed and no similar complaints for this lot number were identified.Corrective actions are in progress.
|
| |
|
Search Alerts/Recalls
|