• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA REVERSE SHOULDER SYSTEM LAT. GLENOSPHERE 36XØ24.5; SHOULDER LATERALIZED GLENOSPHERE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA REVERSE SHOULDER SYSTEM LAT. GLENOSPHERE 36XØ24.5; SHOULDER LATERALIZED GLENOSPHERE Back to Search Results
Model Number 04.01.0207
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Pain (1994)
Date of Event 04/04/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on (b)(6) 2023.Lot 2117351: (b)(4) items manufactured and released on 22-feb-2022.Expiration date: 2027-feb-13.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.Clinical evaluation performed by medical affairs department: revision 8 months after the primary rsa surgery due to glenosphere disassociation from the baseplate.According to report, the glenosphere screw was found intact during the revision.Three polyaxial locking screws were used during the primary surgery; during the revision surgery, the surgeon noted that the locking screws were not locked.The surgeon revised the metaphysis, liner and glenosphere.The surgery was completed successfully.The causes for disassociation cannot be determined with certainty with the elements at hand.Normally, this happens if the taper spigot of the baseplate could not engage fully the glenosphere matching taper.Among the possible reasons for such an occurrence may be a baseplate screw head protruding from the baseplate surface, and preventing the glenosphere to reach its final seating.Additional involved implant: batch review performed on (b)(6) 2023 on reverse shoulder system 04.01.0191 threaded glenoid baseplate ø24.5x30 (k171058 ) lot.2006584 lot 2006584: (b)(4) items manufactured and released on 06-nov-2020.Expiration date: 2025-oct-06.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported event during the period of review.
 
Event or Problem Description
Revision surgery performed 8 months after the primary surgery due to glenosphere disassociation from baseplate (the glenosphere screw was found intact during the revision).Three polyaxial locking screws were used during the primary surgery; during the revision surgery, the surgeon noted that the locking screws were not locked.The surgeon revised the metaphysis, liner and glenosphere.The surgery was completed successfully.It is unknown if the glenosphere guide was used during the primary surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVERSE SHOULDER SYSTEM LAT. GLENOSPHERE 36XØ24.5
Common Device Name
SHOULDER LATERALIZED GLENOSPHERE
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key16813632
Report Number3005180920-2023-00310
Device Sequence Number5691880
Product Code HSD
UDI-Device Identifier07630040728065
UDI-Public07630040728065
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K193175
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number04.01.0207
Device Catalogue Number04.01.0207
Device Lot Number2117351
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2023
Initial Report FDA Received Date04/26/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-