|
EIT EMERGING IMPLANT TECHNOLOGIES GMBH TRIAL, H 8MM, 8°, 32/12; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
|
Back to Search Results |
|
| Model Number |
TFT80832 |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Device report from synthes reports an event in switzerland as follows: it was reported that on (b)(6) 2023, there were problems with the fixation of the conduit tlif trial and implant: with minimal effort, a rotation of the trial or implant can be triggered, which is unsatisfactory for the insertion or positioning of the trial and implant.With a modest amount of effort, one could trigger a rotation of the test cage.This report involves one trial, h 8mm, 8°, 32/12.This is report 12 of 12 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j employee.H3, h6: the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was further reported that the patient was not affected by the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h6: the device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that there was no damage or defects with the trial h 8mm 8deg 32/12.A functional test was not able to be performed as not all mating devices were returned.A dimensional inspection was not performed since it was not applicable to the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the trial h 8mm 8deg 32/12 was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: the following source controlled drawings reflecting the current and manufactured revisions were reviewed: tftxyyzz rev.B (current and manufactured).H4.Part: tft80832, lot: e22di0068, supplier: (b)(4), batch1: lot units were released on 21 oct 2022 with no discrepancies.No non conformance reports were generated during production.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: part:tft81132 lot:e22di0068 supplier: eit - batch1: lot qty of (b)(4) were released on 21 oct 2022 with no discrepancies.No ncrs were generated during production.The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned photos.Visual analysis of the photo revealed that there was no damage or defects with the trial h 8mm 8deg 32/12.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the trial h 8mm 8deg 32/12 was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|