• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. MIC GJ FEEDING TUBE KIT W ENFIT CONNECTOR - ENDOSCOPIC/RADIOLOGIC PLACEMENT; DH EF BALLOON TUBES PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. MIC GJ FEEDING TUBE KIT W ENFIT CONNECTOR - ENDOSCOPIC/RADIOLOGIC PLACEMENT; DH EF BALLOON TUBES PRODUCTS Back to Search Results
Model Number 8250-18
Medical Device Problem Code Misconnection (1399)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.The device history record for lot 30153030 was reviewed and the product was produced according to product specifications.All information reasonably known as of (b)(6) 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.H3 other text : device not returned.
 
Event or Problem Description
Avanos medical inc.Received a single report that referenced two different incidences, which were associated with separate units, involving two different patients.This is the first of two reports.Refer to 9611594-2023-00058 for the second report.Fda medwatch / fda user facility report mw report mw5115942 reported, ¿we have had two instances of patients with gastric-jejuna! tubes in place that have been fed/received medications into the wrong location due to changes in the product.The product previously had the jejunal port as the straight port and the gastric port as the side port.In a recent change by the manufacturer, the port locations have been swapped.This change has been further complicated by the rubbing off of the text that labels the ports as either gastric or jejunai.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIC GJ FEEDING TUBE KIT W ENFIT CONNECTOR - ENDOSCOPIC/RADIOLOGIC PLACEMENT
Common Device Name
DH EF BALLOON TUBES PRODUCTS
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp 84048
MX   84048
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key16815514
Report Number9611594-2023-00057
Device Sequence Number17566437
Product Code KNT
UDI-Device Identifier00350770445458
UDI-Public00350770445458
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K921370
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/06/2023
Device Model Number8250-18
Device Lot Number30153030
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2023
Initial Report FDA Received Date04/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-