Model Number MOSAIQ
Medical Device Problem Code Computer Software Problem (1112)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/29/2023
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that mosaiq did not record the therapy.
 
Additional Manufacturer Narrative
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
 
Additional Manufacturer Narrative
The investigation was completed by conducting a thorough evaluation of the product and the reported information.The customer reported that mosaiq did not record the treatment.The investigation found that the software logs indicate that there was no treatment delivered.From the machine log files it was ascertained that the prescribed dose was delivered as intended.The customer manually recorded the delivered treatment.For an unknown reason, mosaiq never received data from the linac (machine) indicating that the treatment was happening.Mosaiq can only record data that is received from the machine.Based on the information provided there was no mistreatment.Mosiaq did not have a malfunction and was working as designed and intended.
 
Event or Problem Description
The customer reported that mosaiq did not record the treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIQ
Common Device Name
ACCELERATOR, LINEAR, MEDICAL
Manufacturer (Section D)
ELEKTA SOLUTIONS AB
kungstensgatan 18
box 7593
stockholm, SE-10 3 93
SW  SE-103 93
Manufacturer (Section G)
ELEKTA SOLUTIONS AB
400 perimeter center terrace
suite 50
atlanta GA 30346
Manufacturer Contact
cornerstone
london road
crawley, west sussex RH10 -9BL
MDR Report Key16815671
Report Number3015232217-2023-00024
Device Sequence Number17563495
Product Code IYE
UDI-Device Identifier00858164002282
UDI-Public(01)00858164002282(10)2.81.070
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K183034
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberMOSAIQ
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2023
Supplement Date Received by Manufacturer03/30/2023
Initial Report FDA Received Date04/26/2023
Supplement Report FDA Received Date07/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/03/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1