• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH CHUCK WITH KEY/REAM SPEED FOR TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH CHUCK WITH KEY/REAM SPEED FOR TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 05.001.207
Medical Device Problem Codes Mechanical Jam (2983); Complete Loss of Power (4015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/18/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported from russia that during service and evaluation, it was determined that the chuck with key device color band was chipping/fading, the etching was worn, the device had corrosion/rusting/pitting, was worn, the chuck seized ¿ was frozen/will not move, moving parts of the device did not move smoothly, and the etching was illegible.It was further determined that the device failed pretest for general condition, marking and labeling, check drill chuck with key, check the maximum range of tool coupling, check the minimum range of tool coupling, check pins length of handpiece coupling, check for mechanical free movement, and check general function in running mode.It was noted in the service order that the equipment did not work.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device was frozen/will not move - chuck seized.Therefore, the reported condition that the equipment was not working was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to wear.Udi (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHUCK WITH KEY/REAM SPEED FOR TRS
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
SYNTHES PRODUKTIONS GMBH
hauptstrasse 24
waldenburg 4437
SZ   4437
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key16817383
Report Number8030965-2023-05310
Device Sequence Number1906775
Product Code HWE
UDI-Device Identifier07611819977877
UDI-Public07611819977877
Combination Product (Y/N)N
Initial Reporter CountryRS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number05.001.207
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2023
Initial Report FDA Received Date04/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2009
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-