SYNTHES GMBH CHUCK WITH KEY/REAM SPEED FOR TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
05.001.207 |
| Medical Device Problem Codes |
Mechanical Jam (2983); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the chuck seized - frozen/will not move.Therefore, the reported condition that the equipment was not working was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to wear.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported from russia that during service and evaluation, it was determined that the chuck with key device color band was chipping/fading, was worn, the chuck seized ¿ was frozen/will not move, moving parts of the device did not move smoothly, and the etching was illegible.It was further determined that the device failed pretest for general condition, marking and labeling, check drill chuck with key, check the maximum range of tool coupling, check the minimum range of tool coupling, check for mechanical free movement, and check general function in running mode.It was noted in the service order that the equipment did not work.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|