MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
|
Back to Search Results |
|
| Model Number |
MMT-723NAS |
| Medical Device Problem Codes |
No Display/Image (1183); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Pump monitored for several days.Pump received with normal operating current.The stop (idle) current and run current measurement tests are within specification.Pump passed displacement test and a21 alarm test.No battery power loss alarm during testing.Pump history download using thds was successful however, on (b)(6) 2019 there is no alarm reports event date noted.The following were noted during visual inspection: cracked belt clip slot, cracked battery tube threads and cracked reservoir tube lip.The test p-cap/reservoir does lock into place.No moisture damage found on the electronics, motor, battery tube and vibrator assembly noted.After inserting battery unit power properly.Pump monitored for several days and no blank display noted.No unexpected battery power loss anomalies noted.No unexpected failed battery alarm not confirmed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the customer received a battery-failed alarm and the pump was not turning on.Troubleshooting was performed.The customer states the contacts on the battery cap are not missing or damaged.The customer states the display returned and advised battery change resolved issue and we will ship a courtesy battery cap.No harm requiring medical intervention was reported.It was unknown whether the customer will discontinue the use of the device.The pump was returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Pump monitored for several days.Pump received with normal operating current.The stop (idle) current and run current measurement tests are within specification.Pump passed displacement test and a21 alarm test.No battery power loss alarm during testing.Pump history download using thds was successful however, on (b)(6) 2019 there is no alarm reports event date noted.The following were noted during visual inspection: cracked belt clip slot, cracked battery tube threads and cracked reservoir tube lip.The test p-cap/reservoir does lock into place.No moisture damage found on the electronics, motor, battery tube and vibrator assembly noted.After inserting battery unit power properly.Pump monitored for several days and no blank display noted.No unexpected battery power loss anomalies noted.No unexpected failed battery alarm not confirmed.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|