|
(b)(4).This follow-up report is being submitted to relay additional and/or corrected information.Concomitant medical devices - proximal humerus, left, long, 8.5x220mm, item# (b)(4), lot# 3101637, blunt tip screw 4x40mm item#(b)(4)lot# 3078227, blunt tip screw 44mm item# 47-2486-044-40 lot# 3024704,and torque screwdriver item, item# (b)(4), lot unknown.Report source ¿ foreign ¿ ¿japan".Multiple mdr reports were filed for this event, please see the associated reports: 0009613350-2023-00045, 0009613350-2023-00015, and 0009613350-2023-00194.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device manufacturing records confirmed no abnormalities or deviations that might be related to the reported event.Devices used for treatment.Medical records were not provided.Based on the investigation, it could can be assumed that possible contributing factors contributing to the migration of the screws might could be multifactorial related to either patient condition, behavior or implantation procedure.If and to what extent any of these aspects may have influenced the migration of the screw remains unknown.An additional deeper investigation was performed which identified the design limitation of the locking mechanism of the nail as a potential contributing factor.As part of the deeper investigation, a scientific literature search was conducted and a systematic review of the outcome of intramedullary nailing for acute proximal humerus fracture showed that screw migration and perforation into the joint is the second most common complication following secondary loss of reduction.In addition, scientific literature of competitor and predicate devices were assessed.This review has shown that the proximal humeral system performs better and/or in equal range in comparison with legally marketed similar devices.In conclusion, as the cause may be multifactorial an exact root cause could not be identified for the migration of the screw.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
It was reported that two (2) weeks after initial surgery, the surgeon found one of the proximal screw was backed out from the proper position.Revision surgery is not planned, and the patient's condition will continue to be monitored.Due diligence is in progress for this complaint; to date no additional information or product has been received.
|