• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC 6.5CM PEDIATRIC CRANIOTOME; MOTOR, DRILL, ELECTRIC - CRANIOTOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC 6.5CM PEDIATRIC CRANIOTOME; MOTOR, DRILL, ELECTRIC - CRANIOTOME Back to Search Results
Model Number CRANI-P
Medical Device Problem Codes Device Damaged by Another Device (2915); Complete Loss of Power (4015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was traced to component failure due to normal wear.Udi: (b)(4).
 
Event or Problem Description
It was reported from switzerland that during service and evaluation, it was determined that the craniotome device neuro-tip was bent, the bearing was worn, could not insert cutter and was vibrating.It was noted in the service order that the device did not work at all.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6.5CM PEDIATRIC CRANIOTOME
Common Device Name
MOTOR, DRILL, ELECTRIC - CRANIOTOME
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
3035526892
MDR Report Key16822740
Report Number1045834-2023-00647
Device Sequence Number12477277
Product Code HBC
UDI-Device Identifier00845384001775
UDI-Public00845384001775
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K011444
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCRANI-P
Device Catalogue NumberCRANI-P
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/07/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2023
Initial Report FDA Received Date04/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2013
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-