C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER Back to Search Results
Model Number 0119L20
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the last time this occurred the foley balloon was over inflated, as too large a syringe was used causing extra pressure, this happened three times on the same patient.It was a dual lumen foley for irrigation and it was a yellow rubber material, not silicone like most of their foley catheters.A 10ml syringe was used but only 6 ml sterile water was drawn up in the syringe.The first catheter held water on test but ruptured on use but had only been inflated with approximately 3ml of water when it ruptured.The second foley balloon ruptured when tested and some pressure was applied before using on the patient.The third foley balloon was inspected visually and due to looking cracked or dry-rotted was not even tested based on the other failures.The packaging for the product indicated to inflate with 10ml in the balloon.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Additional Manufacturer Narrative
The reported issue was inconclusive as no sample was returned for evaluation.A potential root cause for this failure could be due to "inadequate valve design".It was unknown whether the device had met specifications.Based on patient code 2645 the product was not used on the patient, but it was unknown whether the product had caused the reported failure.The provided lot number was invalid therefore dhr review can¿t be completed.The instructions for use were found adequate and state the following: "sterile unless package is opened or damaged.Warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage latex and may cause the balloon to burst.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Single patient use only.Do not reuse.Do not resterilize.For urological use only.Valve type: use luer slip tip syringe.Do not use needle.Recommended inflation capacities 3cc balloon: use 5ml sterile water 5cc balloon: use 10ml sterile water 15cc balloon: use 20ml sterile water 20cc balloon: use 25ml sterile water 30cc balloon: use 35ml sterile water 40cc balloon: use 45ml sterile water 75cc balloon: use 80ml sterile water do not exceed recommended capacities.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.Bard and bardex are registered trademarks of c.R.Bard, inc.Or an affiliate.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that the last time this occurred the foley balloon was over inflated, as too large a syringe was used causing extra pressure, this happened three times on the same patient.It was a dual lumen foley for irrigation and it was a yellow rubber material, not silicone like most of their foley catheters.A 10ml syringe was used but only 6 ml sterile water was drawn up in the syringe.The first catheter held water on test but ruptured on use but had only been inflated with approximately 3ml of water when it ruptured.The second foley balloon ruptured when tested and some pressure was applied before using on the patient.The third foley balloon was inspected visually and due to looking cracked or dry-rotted was not even tested based on the other failures.The packaging for the product indicated to inflate with 10ml in the balloon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16823295
Report Number1018233-2023-02925
Device Sequence Number15200499
Product Code EZC
UDI-Device Identifier00801741019074
UDI-Public(01)00801741019074
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K910846
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0119L20
Device Catalogue Number0119L20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/06/2023
Supplement Date Received by Manufacturer09/06/2023
Initial Report FDA Received Date04/27/2023
Supplement Report FDA Received Date09/07/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-