During remote monitoring, the device delivered high voltage therapy in response to ventricular tachycardia/fibrillation, however, the shocks were unable to successfully convert the arrhythmia.When the device delivered the inadequate shocks, low defibrillation impedance was observed on the right ventricular (rv) lead.The patient was later seen in-clinic, and the device was reprogrammed.At a later date, the physician elected to explant and replaced the rv lead.The patient was in stable condition.
|
The reported events of inadequate therapy and low defibrillation impedance were confirmed.As received, a complete lead was returned in one piece.X-ray examination of the lead found one ring electrode conductor cable was separated under the right ventricular shock coil region.An internal short was found between the ring electrode and right ventricular conductor paths.After removing the right ventricle shock coil to examine the condition of the insulation beneath, an internal insulation abrasion was found breaching the ring electrode and right ventricular conductor cable lumens with one ring electrode conductor cable abraded/melted/separated in the abrasion zone.The cause of the reported events was due to the exposure and melted/separated of one ring electrode conductor cable at the abrasion zone beneath the right ventricular shock coil.
|