INSULET CORPORATION OMNIPOD DASH INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN Back to Search Results
Model Number 18320
Device Problem Premature Activation (1484)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/23/2023
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned/received to date.If the device is received, a supplemental report will be submitted with the investigation results.We are unable to confirm the reported needle mechanism failure or to determine its root cause.Lot release records were reviewed and the product lot met all acceptance criteria.Specifically, a pod is paired to a pdm and put through simulated use testing including confirming that the device deployed the cannula correctly.
 
Event Description
It was reported the needle mechanism deployed early.The pod was not worn and no adverse patient impact was reported.
 
Manufacturer Narrative
The device was received undeployed.No timeouts or drive stalls were seen in the download data.The data indicates that the pod completed only the first priming sequence before being deactivated.The device was found to function as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD DASH INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
thom mcnamara
100 nagog park
acton, MA 01720
9786007000
MDR Report Key16824853
MDR Text Key314109390
Report Number3004464228-2023-10852
Device Sequence Number1
Product Code LZG
UDI-Device Identifier20385082000020
UDI-Public(01)20385082000020(11)220602(17)231202(10)PD1K06022221
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K211575
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/02/2023
Device Model Number18320
Device Catalogue NumberBLE-I1-529
Device Lot NumberPD1K06022221
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/26/2023
Initial Date FDA Received04/27/2023
Supplement Dates Manufacturer Received08/17/2023
Supplement Dates FDA Received08/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-