The device was returned to olympus for evaluation, and the customer's allegation was confirmed.Additionally, the device evaluation found the charged coupled device was damaged and the adhesive on the protective coating of the bending portion of the device had a chip.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 5 years since the subject device was manufactured.Based on the results of the investigation, it is likely the cause of the event was due to breakage of the image sensor unit including disconnection by stress of repeated use, external factors, handling, or that the components including integrated circuit (ic) chip and capacitor mounted on the electric circuit board had a defect.However, a definitive root cause cannot be identified.The following information is stated in the instructions for use (ifu) ¿chapter 3 preparation and inspection, 3.8 inspection of the endoscopic system¿: "confirm that the wli and nbi endoscopic images are normal.Before inspection, wipe the objective lens using clean lint-free cloths moistened with saline solution or sterilized water; observe the palm of your hand in the wli and nbi endoscopic images; confirm that light is output from the endoscope¿s distal end; adjust the brightness level as appropriate; confirm that the wli and nbi endoscopic images are free from noise, blur, fog, or other irregularities; turn the angulation control levers slowly in each direction until it stops; confirm that the wli and nbi endoscopic images do not momentarily disappear or display any other irregularities." olympus will continue to monitor the field performance of this device.
|