| Model Number |
USSL-1700260 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/17/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This mdr report is based on the reclassification of a previously reviewed complaint.During preparation for a case in germany, a user found the light fiber of the illuminoss implant broken after taking it out of the package.The device was not used.Another illuminoss implant was used.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr report is based on the reclassification of a previously reviewed complaint.Followup information provided by the user: the user reported that the implant was broken in the package and was not at the table.They directly used a new implant.Records review: a review of manufacturing records was performed, and found that the device met speciations at the time of manufacture and release, including 100% inspection for light output of assembled light fibers.Returned product evaluation: the device was received on 18nov2019 for evaluation.Visual inspection was performed, as well as inspection under magnification.The product defect was verified.The light fiber was observed broken at the y-connector.The product evaluation was not able to identify the cause of the light fiber break.Conclusion: the firm was unable to determine the cause of this reported malfunction.
|
| |
|