| Model Number |
0502503030 |
| Medical Device Problem Code |
Sharp Edges (4013)
|
| Health Effect - Clinical Code |
Abrasion (1689)
|
| Date of Event |
04/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient was scratched by the sharp distal tip.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was scratched by the sharp distal tip.
|
| |
|
Additional Manufacturer Narrative
|
|
This scope was not received for evaluation at stryker endoscopy.This scope was received at henke for evaluation.Based on the henke service record , the reported failure "sharp tip" was confirmed.According to henke: scope evaluated as a level 3.Distal tip and fiber damage, broken lenses, distal cover glass damage.Probable root cause for this failure is: the complaint for ¿sharp tip¿ has been confirmed and is due to customer use and handling.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|