Model Number WA22507D
Medical Device Problem Codes Break (1069); Smoking (1585)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to olympus that an hf-resection electrode loop suffered too much mechanical manipulation and when activated with the hf energy it broke and there was smoke.An endoscopic search was carried out in the uterine cavity and pieces of material were found.The debris was able to be recovered from the uterine cavity and there was no impact to the patient.The procedure was a therapeutic hysteroscopy which was completed using a similar device.This medwatch is being submitted for the reportable event.
 
Additional Manufacturer Narrative
The device was returned to olympus for evaluation and repair.The investigation is still in progress; therefore, the root cause of the reported defect cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.In general, the customer is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.Olympus will continue to monitor the field performance of this device.Additional pma/510(k): k120418.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the cause of the reported issue is most likely due to the application of excessive force.In general, the customer is required to check the function of all devices used prior to a procedure.Additionally, according to the instructions for use, a suitable replacement device must be provided during an application.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP
Common Device Name
ELECTRODE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16828211
Report Number9610773-2023-01164
Device Sequence Number13908550
Product Code FAS
UDI-Device Identifier14042761068236
UDI-Public14042761068236
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K100275
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA22507D
Device Catalogue NumberWA22507D
Device Lot Number1000082810
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/29/2023
Supplement Date Received by Manufacturer10/06/2023
Initial Report FDA Received Date04/27/2023
Supplement Report FDA Received Date10/09/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-