The device was received for evaluation.During functional testing, the loudspeaker did not work.A service history review was performed and revealed that the device has no previous service events; therefore, servicing did not cause or contribute to the reported event.The reported condition was verified.The cause of the condition was determined to be a damaged speaker wire.The rear case requires replacement to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|