| Model Number |
N/A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Capsular Bag Tear (2639)
|
| Date of Event |
03/30/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Measure: this issue only impacts this specific device.Analyze: (b)(6) reviewed the open call for lls5111.Case #(b)(4) - fair centration with adequate suction is noted.The capsulotomy firing starts at frame 3 and completes at frame 10.Fragmentation beginning right after on frame 11 and completing at frame 27.Patient movement is not noted throughout the entirety of the procedure.Laser functioned as designed root cause: surgical technique during to cortex removal.
|
| |
|
Event or Problem Description
|
|
Procedure id (b)(4).During cortex removal after (b)(4) dr.Noted a fragmentation imprint on the cortex.When she pulled the first piece of cortex with ia the vitreous came into the anterior chamber and there appeared to be no posterior capsule.An anterior vitrectomy was performed and a 3 piece iol was placed in the sulcus.
|
| |
|