• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LENSAR, INC LENSAR LASER SYSTEM-FS 3D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LENSAR, INC LENSAR LASER SYSTEM-FS 3D Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Capsular Bag Tear (2639)
Date of Event 03/30/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Measure: this issue only impacts this specific device.Analyze: (b)(6) reviewed the open call for lls5111.Case #(b)(4) - fair centration with adequate suction is noted.The capsulotomy firing starts at frame 3 and completes at frame 10.Fragmentation beginning right after on frame 11 and completing at frame 27.Patient movement is not noted throughout the entirety of the procedure.Laser functioned as designed root cause: surgical technique during to cortex removal.
 
Event or Problem Description
Procedure id (b)(4).During cortex removal after (b)(4) dr.Noted a fragmentation imprint on the cortex.When she pulled the first piece of cortex with ia the vitreous came into the anterior chamber and there appeared to be no posterior capsule.An anterior vitrectomy was performed and a 3 piece iol was placed in the sulcus.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LENSAR LASER SYSTEM-FS 3D
Common Device Name
LENSAR LASER SYSTEM-FS 3D
Manufacturer (Section D)
LENSAR, INC
2800 discovery drive ste.100
orlando FL 32826
Manufacturer (Section G)
LENSAR, INC
2800 discovery drive ste. 100
orlando FL 32826
MDR Report Key16829668
Report Number3009026057-2023-54614
Device Sequence Number1931503
Product Code OOE
UDI-Device Identifier00867744000112
UDI-Public(01)00867744000112(11)200303
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K182795
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number70-00013-001
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2023
Initial Report FDA Received Date04/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/03/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-