Model Number PT-000011
Device Problem Failure to Power Up (1476)
Patient Problem Diabetic Ketoacidosis (2364)
Event Date 04/24/2023
Event Type  Injury  
Event Description
It was reported by the patient's doctor that the patient is currently being hospitalized with diabetic ketoacidosis (dka).The patient's blood glucose levels rose above 250 mg/dl.The personal diabetes manager (pdm) reportedly will not turn on.When the pdm is plugged into the charger, the screen remains blank and there is no charging symbol.The patient is unable to use the pdm for treatment and does not have a backup method for insulin delivery.The patient is currently being treated with an insulin infusion at the hospital.The doctor requested for a new pdm to be sent to the hospital and the patient will be discharged once the pdm is received.
 
Manufacturer Narrative
The device has not been returned/received to date.If the device is received, a supplemental report will be submitted with the investigation results.We are unable to determine if any product condition could have contributed to the reported hospitalization and diabetic ketoacidosis.Lot release records were reviewed and the product lot met all acceptance criteria.Specifically, a pod is paired to a pdm and put through simulated use testing including communicating with the pdm, deployment, delivering fluid, occlusion detection, and freedom from hazard alarms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD DASH INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
thom mcnamara
100 nagog park
acton, MA 01720
9786007000
MDR Report Key16830846
MDR Text Key314159221
Report Number3004464228-2023-10947
Device Sequence Number1
Product Code LZG
UDI-Device Identifier10385082000146
UDI-Public(01)10385082000146(11)221210(10)L000506
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K211575
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 04/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberPT-000011
Device Catalogue NumberINT1-D001-MG
Device Lot NumberL000506
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/25/2023
Initial Date FDA Received04/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/10/2022
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age19 YR
Patient SexFemale
-
-