| Model Number |
PLR3955LT |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Burn(s) (1757)
|
| Date of Event |
04/20/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not yet been made available for evaluation.Should further information or the device become available, a follow-up report will then be issued.
|
| |
|
Event or Problem Description
|
|
The provider alleges the customer alleges that the lift chair has two burn spots on it and the lift chair allegedly burnt his back.
|
| |
|
Event or Problem Description
|
|
The provider alleges the customer alleges that the lift chair has two burn spots on it and the lift chair allegedly burnt his back.
|
| |
|
Additional Manufacturer Narrative
|
|
There were 2 corresponding burn marks at the seat base interface.Given the uni-directional discoloration of the poly-fill underneath the burn spots along with the normal operating temperature of both heat pads, the evaluator believes the burn spots were caused by an outside heat source unrelated to the chair.
|
| |
|