| Model Number |
MT22949 |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that no vibration occurred.No product or data was provided for evaluation.The allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The product was evaluated.An external visual inspection was performed and passed.Receiver charge and boot tests were performed and passed.Functional tests were performed and passed.A manual vibration test was performed and passed.The vibrator motor resistance was measured and passed.An internal visual inspection was performed and passed.The receiver log was downloaded and reviewed finding no error related to the complaint.The allegation was not confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|