SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/24/2023
Event Type  malfunction  
Event Description
The customer reported to olympus, the high flow insufflation unit does not stay on because the power switch does not stay in place when pressed.The issue was found during preparation for use.The intended ear/nose/throat procedure was completed using a similar device.There were no reports of patient harm.
 
Manufacturer Narrative
To date, the device has not been returned to olympus for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The customer sent the device to a third party repair center.The reported fault, "unit does not stay on because power switch does not stay in place when pressed", was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the power would not turn on even if the power switch was pressed due to a failure of the power switch.The definitive root cause of the damaged power switch could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16838304
MDR Text Key314473519
Report Number3002808148-2023-04383
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-