MAUDE Adverse Event Report: CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 SYS PED.WITH PRECHAMBER; HYDROCEPHALUS MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 SYS PED.WITH PRECHAMBER; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX438T
Medical Device Problem Codes Mechanical Problem (1384); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 03/30/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that a progav 2.0 (#fx428t) was implanted during a procedure performed on an unknown date.According to the complainant, the valve showed adjustment difficulties.The patient underwent a revision procedure.The complaint device will be return to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.
 
Event or Problem Description
It was reported that a progav 2.0 (#fx438t) was implanted during a procedure performed on (b)(6) 2021.According to the complainant, the shunt showed an over-drainage and adjustment difficulties.The patient underwent a revision procedure performed on (b)(6) 2023.The complaint device has returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 2 years, 7 months.Weight: 14 kilograms (kg).Height: 93.5 centimeters (cm).Gender: male.
 
Additional Manufacturer Narrative
Visual inspection: in the first step of our investigation, we performed a visual inspection of the product.We have checked for possible damage, deformation of the housing, deposits or other abnormalities.The following observations were made during the visual inspection: scratches on the outer housing permeability test: the permeability test was performed at a hydrostatic pressure difference of the pressure setting of the gav® 2.0 upon receipt plus 30 cmh2o in the horizontal direction of flow.The test showed that the progav 2.0 with the pediatric prechamber is permeable.Computer control test: to check the flow rate of the valve, the valve was tested on a miethke computer controlled testing apparatus and passed the standard test.The flow of the test liquid was reduced step by step from 60 ml/h to 5 ml/h (in accordance with iso 7197).Distilled water was used as the test-liquid.The resulting pressure was measured.The computer controlled test showed the opening pressure of the progav 2.0, at a reference flow rate of 20 ml/h in a horizontal position, to be 3.97 cmh2o.This is not within the specified tolerance of 12.5 cmh2o ± 3 cmh2o.Additional testing confirms the first test result.According to our results, we can detect a presence of a slowed accelerated outflow.Adjustability test: in this step, it was investigated whether the adjustable progav 2.0 can be successfully set to each specified pressure setting.It was checked whether the valve is fully adjustable within the full range of specified pressure settings ( in increments of 5 cmh2o).The progav 2.0 was found to be non-adjustable within the specified range.Braking force and brake function test: the braking force and brake function test investigates whether the brake function of the adjustable valves is present and how much force must be exerted on the housing to release the rotor to adjust the valves using the integrated magnet of a specific measurement apparatus of braking force.The braking force test indicates that the brake function of the progav 2.0 is present.Due to the non-adjustability of the valve, as detailed in section 7.4, an investigation of the braking force was not possible.Internal inspection of product: in order to verify whether the investigated valve was compromised by the known risks of hydrocephalus therapy, e.G.By a build-up of natural substances (protein, blood, or tissue particles) in the cerebrospinal fluid, the valve is finally opened with a special tool.After dismantling of the valve, deposits were found in progav 2.0.For more detailed visibility, the proteins/deposits in progav 2.0 were coloured by using a colouring solution.[colouring solution consisting of coomassi brilliant blue r mixed with 0.6% ethanol and distilled water].Results: based on our investigation results, we have determined that there was an accelerated outflow and a non-adjustability in the valve at the time of our investigation.Detected deposits can be the cause of the malfunction existing deposits in the csf can temporarily impair the function of the valve and is described as concomitant symptoms in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.Further actions: from our point of view, no further regulatory actions are required.Return of product: we will return the investigated product to the sender for our own relief.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 SYS PED.WITH PRECHAMBER
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key16838793
Report Number3004721439-2023-00109
Device Sequence Number20539049
Product Code JXG
UDI-Device Identifier04041906137434
UDI-Public4041906137434
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/12/2026
Device Model NumberFX438T
Device Catalogue NumberFX438T
Device Lot Number20051395
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2023
Supplement Date Received by Manufacturer06/30/2023
Initial Report FDA Received Date04/28/2023
Supplement Report FDA Received Date07/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age31 MO
Patient SexMale