Model Number 3501
Device Problems Signal Artifact/Noise (1036); Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
Patient Problem Electric Shock (2554)
Event Date 04/11/2023
Event Type  Injury  
Manufacturer Narrative
No additional information is available.If additional information becomes available the report will be updated at that time.
 
Event Description
It was reported that the subcutaneous implantable cardioverter defibrillator (s-icd) exhibited oversensing of noise leading to three inappropriate shocks.The patient was brought into the office and the sensing vector was reprogrammed from primary to secondary.The data was sent to technical services (ts) for review.Upon review, ts noted that the s-icd recorded three treated and one untreated episode since being implanted.The untreated episode was eight months earlier.The three treated were all in the current month.All episodes were due to oversensing of noise and baseline shifting.Ts discussed that the noise is due to sudden changes in the impedance pathway of the sensing vector.Ts further discussed possible causes and other troubleshooting options.The s-icd and electrode remain in service and no adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM S-ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16838986
MDR Text Key314260253
Report Number2124215-2023-20727
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P110042/S077
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/28/2022
Device Model Number3501
Device Catalogue Number3501
Device Lot Number158051
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/11/2023
Initial Date FDA Received04/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/28/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-