| Model Number |
1804250-18 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
It was reported that during preparation and before use, when the stent delivery system was removed from the package, the shaft was observed to have separated in two pieces.There was no damage reported to the device packaging.The device was not used.There was no patient involvement.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported shaft separation was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation was unable to determine a conclusive cause for the reported shaft separation.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|