| Model Number |
V11460-010 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device history record review was completed on the reported lot v2s056.The lot was found to have been manufactured and released to predetermined specifications.No anomalies were found during review of the records.We received two samples for investigation that were not already activated.One sample was determined to have an incomplete top seal due to machine setup.The root cause was determined to be a burst pack due to an incomplete top seal.The machine was setup correctly and is designed with a mechanism which reduces the likelihood for this issue.Cardinal health will continue to monitor complaint trends for this reported issue.
|
| |
|
Event or Problem Description
|
|
Customer reported nurse activated novaplus+ infant heel warmer as usual and the warmer opened and the content made contact with her tee shirt.Nurse stated she did not feel any splash onto her face but as a precaution i suggested she rinse her eyes with normal saline/water since some contents came in contact with her eye brow.No other demographics provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial report was filled on product code 11460-010.Based on the lot reported, the correct product number is v11460-010.Cardinal health has received an increase in burst/leaking complaints regarding the novalplus infant heel warmer with tape.Customer complaints have noted rupturing and bursting of the device when activation is attempted, making the device unusable and in many cases causing the contents inside the pouch to leak and contact clinicians, patients, and other individuals or surfaces in proximity.Samples returned to cardinal health from customers have been evidenced to have an incomplete top seal, resulting in the malfunction.There have been no reportable adverse events/injuries associated with v11460-010, lot v2s056.Cardinal health has halted distribution of the novalplus infant heel warmer w/tape for lot number v2s056 and initiated a voluntary recall on june 16, 2023.
|
| |
|
Search Alerts/Recalls
|