MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PLUS PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Model Number |
3502750 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
04/12/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: as of now there is no information regarding the explant or reoperation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a 55-year-old female patient underwent revision breast augmentation with a 750cc mentor smooth round moderate plus profile and experienced breast implant deflation on her right side postoperatively.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.
|
| |
|
Additional Manufacturer Narrative
|
|
On may 19, 2023, the mentor failure analysis lab received the device for evaluation.As per device and paperwork received, the following information has been updated on this form: - is required intervention has been selected under section b; field b2 - fields d6b for explantation date have been updated to (b)(6) 2023.- field h6 health effect - impact code ¿device explantation¿ has been added on may 26, 2023, device evaluation was completed.Device evaluation summary: mentor conducted visual inspection of the returned device.Visual analysis of the returned sample revealed that the sm mpps dv 750cc breast implant had a crease/fold on the posterior view.In addition, a tear was found within the crease/fold, measuring approximately 11 cm.The evaluation determined that the cause of the rupture is consistent with normal wear.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in a deflation in a later time.Breast implants may also simply wear out over time.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Capa activity was not required.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On september 22, 2023, mentor became aware that manufacturer report number 1645337-2023-04681, is a duplicate of manufacturer report number 1645337-2023-06752 and and all reporting will continue in this file manufacturer report number 1645337-2023-04681.On october 11, 2023, mentor became aware that patient received the following replacement devices on (b)(6) 2023: right side: catalog number 3502800; serial number (b)(6), and left side: catalog number 3502800; serial number (b)(6).Manufacturer¿s reference number: (b)(4).
|
| |
|