| Catalog Number |
02-010-06-0220 |
| Medical Device Problem Codes |
Fitting Problem (2183); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994)
|
| Date of Event |
01/01/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported via clinical study, that the 65 yo female patient experienced aseptic femoral loosening with persistent thigh pain related to micromotion of femoral stem with incomplete fixation.The date of adverse event onset is unk-unk-2016.Initially tried strengthening exercises as patient treatment.The patient was revised on (b)(6) 2021 along with medication.The patient¿s outcome was last known as resolved on (b)(6) 2021.
|
| |
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 02-010-06-0220 - 4114527 - logic cc femoral size 2, left.02-012-45-2020 - lgc tibial fit tray cem sz 2f / 2t.02-012-60-1680 - logic stem ext 16mm x 80mm.208-05-02 - cc distal fem augment sz 2, 5mm.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported cannot be confirmed from the information provided but may be the result of femoral loosening.The reported femoral loosening could not be confirmed.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|